Field Fox Calibration and Testing
Note: text indicates physical buttons and text indicates an on-screen option.
Note: Start and Stop Frequencies must be set separately for both IL and VSWR.
Calibrate for Insertion Loss
- Mode › Power Meter USB This is the mode for measuring Insertion Loss.
- Freq/Dist › Src Start › Enter start frequency › GHz or MHz
- Freq/Dist › Src Stop › Enter stop frequency › GHz or MHz
- 1 Measure › Src Power Measure
- Connect test cable to Field Fox Port 1. Connect Power Sensor USB to Field Fox USB port.
- Connect Power Sensor N-Type Connector to the end of the test cable, adding any adapters that will be used
for the circuit under test.
› Measure › Let the marker run across the graph until you are returned to the previous options.› Src Data → Mem › Done This saves the calibration data for the power sensor.
- 1 Measure › Gain This sets the Field Fox to measure dB Gain/Loss The next steps check accuracy of calibration.
- Scale/Amptd › Autoscale › Let the marker run across the graph one complete time.
- Run/Hold › “Hold” will appear in red text on the left of the screen, indicating that the Field Fox is no longer taking measurements.
- Scale/Amptd › Autoscale
- Mkr/Tools › Minimum › IF the minimum is less (more negative) than -0.1 (upper right of the screen), then ensure physical connections are tight, and restart insertion loss calibration steps from 1 Measure › Src Power Measure. Otherwise, if greater (more positive) than -0.1, continue. A number closer to 0 is better.
- NOTE: Sometimes the Field Fox will not automatically find the actual minimum from Mkr/Tools. If this happens, press Marker › Use the physical wheel to locate the real minimum.
Calibrate for VSWR
- Mode › CAT/TDR This is the Mode for measuring VSWR.
- 1 Measure › VSWR
- Freq/Dist › Src Start › Enter start frequency › GHz or MHz
- Freq/Dist › Src Stop › Enter stop frequency › GHz or MHz
- 5 Cal › Mechanical Cal
› Change DUT Connectors › Select Type N› If necessary, Change Gender to match Cal Kit.› Next › Select 85518A or 85515A or whichever Cal Kit you are going to use.› Finish› Start Calibration › Follow instructions on screen.
- Connect test cable to Field Fox Port 1.
- Connect Cal Kit OPEN to test cable, adding any adapters that will be used for the circuit under test. › Measure
- Change the Cal Kit OPEN to the Cal Kit SHORT › Measure
- Change the Cal Kit SHORT to the Cal Kit LOAD › Measure
- Finish
- 1 Measure › VSWR The next steps check accuracy of calibration.
- Scale/Amptd › Autoscale
- Run/Hold › “Hold” will appear in red text on the left of the screen, indicating that the Field Fox is no longer taking measurements.
- Scale/Amptd › Autoscale
- Mkr/Tools › Peak › IF the peak is greater than 1.019 (upper right of the screen), then ensure physical connections are tight, and restart VSWR calibration steps from 5 Cal. Otherwise, if less than 1.02, continue. A number closer to 1 is better.
Set Title for Insertion Loss
- 7 System › Display › Title › Enter the Title of the graph. For example, “DDG129 R-RA123_IL” or if measuring at multiple frequency ranges, “DDG129 R-RA123_IL 1.0G-2.0G”
Measure Insertion Loss
- Mode › Power Meter USB This is the mode for measuring Insertion Loss.
- Connect test cable to Field Fox Port 1. Connect other end of test cable to circuit under test.
- Connect local Range Extender to Field Fox USB port using USB cable.
- Connect Power Sensor USB to distant Range Extender USB port.
- The Range Extenders should be connected to each other via a CAT5 cable.
- Connect Power Sensor N-Type Connector to the distant end of the circuit under test.
- 1 Measure › Gain This sets the Field Fox to measure dB Gain/Loss
- Run/Hold › “Hold” will no longer appear in red text on the left of the screen. If it does appear, press Run/Hold again.
- Scale/Amptd › Autoscale › Let the marker run across the graph one complete time.
- Run/Hold › “Hold” will appear in red text on the left of the screen, indicating that the Field Fox is no longer taking measurements.
- Scale/Amptd › Autoscale
- Mkr/Tools › Minimum NOTE: Sometimes the Field Fox will not automatically find the actual minimum from Mkr/Tools. If this happens, press Marker › Use the physical wheel to locate the real minimum.
Save Graph for Insertion Loss
- NOTE: Ensure Frequency Range and Title and Minimum Marker are correct before saving.
- 9 Save/Recall › File Type › Trace+State
› Save › Enter the same text as the Title › Done
- 9 Save/Recall › File Type › Picture (PNG)
› Save › Enter the same text as the Title › Done
Set Title for VSWR
- 7 System › Display › Title › Enter the Title of the graph. For example, “DDG129 R-RA123_VSWR” or if measuring at multiple frequency ranges, “DDG129 R-RA123_VSWR 1.0G-2.0G”
Measure VSWR
- Mode › CAT/TDR This is the mode for measuring VSWR.
- Connect test cable to Field Fox Port 1. Connect other end of test cable to circuit under test.
- Connect Cal Kit LOAD to the distant end of the circuit under test.
- 1 Measure › VSWR
- Run/Hold › “Hold” will no longer appear in red text on the left of the screen. If it does appear, press Run/Hold again.
- Scale/Amptd › Autoscale › Let the Field Fox take measurements for 5 seconds.
- Run/Hold › “Hold” will appear in red text on the left of the screen, indicating that the Field Fox is no longer taking measurements.
- Scale/Amptd › Autoscale
- Mkr/Tools › Peak
Save Graph for VSWR
- NOTE: Ensure Frequency Range and Title and Peak Marker are correct before saving.
- 9 Save/Recall › File Type › Trace+State
› Save › Enter the same text as the Title › Done
- 9 Save/Recall › File Type › Picture (PNG)
› Save › Enter the same text as the Title › Done